Descripción
Dimensions (W / D / H) 137 x 51 x 98 cm; bench top
Weight 171 kg
Power Requirements 100?120/220?240 VAC, 50/60 Hz
Valve pressure At least 2 bar above process pressure
Flow rate range: 1 mm valve block: up to 40 ml/min; 1.6 mm valve block: up to 100 ml/min; 2.4 mm valve block: up to 300 ml/min
Inlets 6
Outlets 6
Columns 1-8
Maximum operating pressure 270 psi (18.6 bar)
UV sensors 4, LED, dual wavelengths
Conductivity sensors 4, range 0?200 mS/cm
pH sensors 4, range 0?14
Pumps 6, stepper motor direct drive dual piston, PEEK? fluid path; self-flush pistons; accuracy +/? 3%; available flow rate ranges 0?12, 0?36, 0?100, 0?300 ml/min
Valve Block 104 pneumatic two-way valves; interchangeable
Sample injector 1; 6-port, two position, automatic actuation
External outputs 5, synched to switch, cycle, phase, method, and peak collect
Control system ProComposer? (Windows PC; method authoring); BIOController? (Windows PC)